Accuracy of quantitative 177Lu SPECT/CT imaging: A systematic review
Julian C J van Oorschodt1,2, Guy D Eslick3, Martin Ugander1
1Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
None:
Accurate quantification of lutetium-177 (177Lu) activity in SPECT/CT imaging is essential for a further development of 177Lu-based peptide receptor radionuclide therapy (PRRT) in establishing absorbed dose-effect relationships. This review summarizes the reported accuracy of quantitative 177Lu SPECT/CT imaging and highlights key sources of variability, including partial-volume effect (PVE) in small volumes. A systematic review was conducted according to PRISMA guidelines. MEDLINE, EMBASE, and Web of Science databases were searched with no language restrictions. Original studies explicitly reporting on the accuracy of 177Lu activity quantifications in physical SPECT/CT measurements were included. Reported percentage error/deviation and recovery coefficient (RC) were extracted and compiled, with studies grouped by object type (geometric phantoms, anthropomorphic phantoms, and patient data). The literature search identified 630 records, of which 46 studies were included for analysis. Percentage errors of quantifications were found to constitute a large range (-102% to 285%). The recovery of 177Lu activity from small volumes is inherently limited by PVE. The application of PVC has led to improvements in the accuracy and precision of quantifications on small volumes in phantom imaging. The accuracy of 177Lu activity quantifications in SPECT/CT imaging remains heterogenous across literature, is subject to large variability and will be degraded by PVE when imaging small volumes. The data suggests that the implementation of standardized procedures including standardization of calibration, reconstruction, segmentation, and PVC assumptions may lead to an improved accuracy and precision of quantitative 177Lu SPECT/CT imaging in clinical practice, thereby allowing the further development of 177Lu-based PRRT.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT