Evaluation of Parameters Determining the Pull-Out Strength of Poly(L-Lactide-Co-ε-Caprolactone) Barbed Suspension
Salvatore Fundaro1, Dennis Malvin Hernandez Malgapo2,3, Martina Manni4
1Cosmetic Surgery, Qute Clinic, Modena, Italy.
Purpose:
This study aimed to examine a pull-out strength test methodology and compare the pull-out strength of different insertion methods for poly(L-lactide-co-ε-caprolactone) (PLCL) barbed suspension threads.
Methods:
A custom-designed "pull-out strength test device" was used to measure the maximum force required to pull out threads inserted into an ex vivo turkey breast model.
Results:
Insertion techniques influence pull-out strength. Additionally, trend analysis indicated that thinner needles (0.85 mm) generally yielded higher pull-out resistance than larger diameter needles (1.28 mm), suggesting they may create a tighter initial grip within the tissue.
Conclusion:
Different insertion techniques influence the pull-out strength of PLCL in an ex vivo model.