Helium Saver Mode With Nitrogen Make-Up Cooling Gas for GC-MS With Cold EI
Aviv Amirav1,2
1School of Chemistry, Tel Aviv University, Tel Aviv, Israel.
Abstract:
Cold EI uses helium as the column carrier and cooling make-up gases at ~50 mL/min. In the rare case that helium supply could be temporarily interrupted, nitrogen or hydrogen can be used, but hydrogen leads to decomposition at the GC injector, and thus, nitrogen is preferred. However, nitrogen leads to longer analysis time and/or reduced separation. We describe the use of helium as the column carrier gas and nitrogen as the cooling make-up gas. In this "Helium Saver" mode, the helium consumption is as in standard EI and nitrogen can be used alone in stand-by mode for further helium saving.
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Noble Gases
The elements in group 18 are noble gases (helium, neon, argon, krypton, xenon, and radon). They earned the name “noble” because they were assumed to be nonreactive since they have filled valence shells. In 1962, Dr. Neil Bartlett at the University of British Columbia proved this assumption to be false.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...