Investigation of a Low-cost Portable Inline Dynamometer for Measuring Isometric Knee Extensor Strength
Amantej Sran1, Elira Leake2, Sydney Yott1
1College of Engineering University of Guelph.
International Journal of Sports Physical Therapy
|July 6, 2026
Summary
Low-cost inline force sensors show promise for measuring isometric knee extension torque, but individual results require cautious interpretation due to potential measurement errors.
Area of Science:
- Biomechanics and Kinesiology
- Rehabilitation Engineering
- Sports Science
Background:
- Isometric strength testing is crucial for assessing deficits and monitoring progress in rehabilitation and training.
- The accuracy of low-cost inline force sensors compared to expensive lab-based dynamometers remains unclear.
Purpose of the Study:
- To evaluate the agreement between a lab-based dynamometer and low-cost inline force sensors (wireless and wired) for measuring isometric knee extension torque.
Main Methods:
- Prospective cross-sectional study involving 14 healthy adults.
- Measurements of peak isometric knee extension torque at 90°, 45°, and 15° using a Humac dynamometer, a wireless Tindeq sensor, and a wired MLP sensor.
- Agreement assessed via Pearson's Correlation, ICC, MDC, and Bland-Altman LOA; voluntary activation (VA) quantified using electrical stimulation.
Main Results:
- The Tindeq wireless sensor demonstrated minimal bias compared to the Humac dynamometer across knee angles.
- Bland-Altman Limits of Agreement (LOA) were relatively large, indicating potential for significant individual measurement errors.
- Inline sensors (Tindeq and MLP) showed high agreement with each other (ICC > 0.94) but lower agreement with the lab-based dynamometer (ICC < 0.80).
Conclusions:
- Low-cost wireless force sensors like Tindeq show potential for clinical applications in strength assessment.
- Despite small bias, large LOA necessitate cautious interpretation of individual measurements from these devices.
- Further validation with larger sample sizes and repeated testing sessions is recommended.