EVALUATION OF A NOVEL POINT-OF-CARE VISCOELASTIC COAGULATION MONITOR WITH COMPARISON TO THROMBOELASTOGRAPHY AND

Lauren E Mumm1,2, Alex M Lynch2, Laura K Culbreth2

  • 1North Carolina Zoo, Asheboro, NC 27205, USA, lmumm@stlzoo.org.

Insights

This study introduces a portable coagulation analyzer for chimpanzees, establishing crucial reference values for clot formation and strength. Findings reveal sex and age influence coagulation, aiding improved veterinary care.

Area of Science:

  • Veterinary Medicine
  • Primate Hematology
  • Coagulation Diagnostics

Background:

  • Cardiovascular and neoplastic diseases are prevalent in managed chimpanzees, potentially causing hemostatic issues.
  • Coagulation assessments are underutilized in chimpanzee medicine.
  • Portable, point-of-care devices offer advantages over lab-based assays in zoological settings.

Purpose of the Study:

  • To validate the VCM Vet (VCM) device for chimpanzee coagulation analysis.
  • To establish reference values for VCM clot formation and strength parameters.
  • To compare VCM results with traditional thromboelastography (TEG) and conventional coagulation parameter (CCP) assays.

Main Methods:

  • Whole blood samples were collected from anesthetized chimpanzees (n=26).
  • The VCM device was used to analyze coagulation kinetics.
  • Plasma-based TEG and CCP assays were also performed for comparison.
  • Statistical analyses compared VCM profiles across sex, age, and sample characteristics.

Main Results:

  • The VCM device successfully analyzed samples from 25 chimpanzees, establishing interpretable kinetic tracings.
  • Reference values for VCM parameters were determined from 21 healthy chimpanzees.
  • Chimpanzees exhibited fast clot formation and strong clots compared to domestic animals.
  • Females formed stronger clots than males; older chimpanzees (>30 yr) formed clots faster than younger ones (≤30 yr).
  • Baseline values for TEG and CCP were reported for 19 healthy chimpanzees.

Conclusions:

  • The VCM device is a viable tool for assessing chimpanzee coagulation in field or zoo settings.
  • Established reference values provide a foundation for improved coagulation monitoring in managed chimpanzees.
  • Coagulation assessments, including VCM, TEG, and CCP, may offer comparable clinical insights.
  • This study supports enhanced standard of care for chimpanzees through better coagulation assessment.