Safety of endoclips for magnetic resonance imaging
Wingchi Edmund Kwok1, Adam Gentile1, Randy Crane1
1Department of Imaging Sciences, University of Rochester Medical Center, Rochester, New York, USA.
Background:
A fatal incident in which magnetic resonance (MR) may be involved and the appearance of strong MR attraction experienced by some MR-conditional endoclips have cast confusion over the MR safety of MR-conditional endoclips, which may lead to unnecessary delay or denial of MRI to some patients. On the other hand, there are endoclips that are MRI-unsafe, yet not all MRI facilities screen patients for endoclips.
Purpose:
Our purpose is to promote a better understanding of the MRI risks associated with endoclips to facilitate proper management of MRI exams involving patients with endoclips.
Methods:
Deflection angle measurements were conducted on three models of MR-conditional endoclips on a 1.5T and a 3T MRI system to assess the magnetic forces on the endoclips. Detachment force measurements were also performed on porcine stomach mucosa specimens to evaluate the attachment force. Besides, detachment tests were conducted inside the magnet bore to evaluate the effect of magnetic-induced torque.
Results:
The measured detachment force was much larger than the magnetic force for each clip tested, with the ratios of the two forces being 87.3-251 at 1.5T and 46.7-130 at 3T, indicating low detachment risk for these clips. This was further supported by the detachment tests inside the magnet bore in which no endoclip detached.
Conclusions:
This study reinforces the MR conditional labeling on endoclips, demonstrating that the appearance of strong magnetic force does not necessarily imply an endoclip is MR unsafe. If the counter force (i.e., the detachment force) provided by an endoclip is much larger than the MRI magnetic force, as observed in this study, the risk of clip detachment during MRI is low. This paper should assist in facilitating proper management of MRI exams involving patients with endoclips, preventing unnecessary denial of MRI examinations to patients while protecting patient safety.
Insights
MR-conditional endoclips are safe for MRI scans. Despite magnetic attraction, their strong attachment force minimizes detachment risk, ensuring patient safety during magnetic resonance imaging.
Area of Science:
- Medical Devices
- Magnetic Resonance Imaging
- Biomaterials
Background:
- Concerns exist regarding the MRI safety of MR-conditional endoclips due to magnetic attraction.
- A fatal incident and observed magnetic attraction have caused confusion about MR safety.
- Some MRI facilities do not screen for endoclips, posing a risk to patients with MRI-unsafe devices.
Purpose of the Study:
- To clarify MRI risks associated with endoclips.
- To promote better understanding for proper MRI exam management in patients with endoclips.
Main Methods:
- Assessed magnetic forces on MR-conditional endoclips using deflection angle measurements at 1.5T and 3T.
- Measured detachment forces on porcine stomach mucosa to evaluate attachment strength.
- Conducted in-bore detachment tests to assess magnetic-induced torque effects.
Main Results:
- Detachment forces significantly exceeded magnetic forces across all tested clips (ratios 87.3–251 at 1.5T, 46.7–130 at 3T).
- These force ratios indicate a low risk of endoclip detachment during MRI.
- In-bore tests confirmed no endoclip detachment occurred.
Conclusions:
- The study supports the MR conditional labeling of endoclips.
- Strong magnetic force does not automatically mean an endoclip is MR unsafe.
- Low detachment risk is confirmed when endoclip detachment force surpasses MRI magnetic force, aiding safe MRI management.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging