Microbiological profile of hemodialysis water treatment systems from a high-volume center
Magaly Padilla-Orozco1, Samantha Pérez-Cavazos2, Alfredo Castillo-Mariscal3
1Head Hospital Epidemiology, Hospital Universitario "Dr. José Eleuterio González", Universidad Autónoma de Nuevo León, Avenida Francisco I. Madero, Mitras Centro, Monterrey, CP 64460, Mexico.
Purpose:
To document the microbiological isolates in treated water samples obtained from a hemodialysis center and investigate potential correlations between these isolates and the microorganisms identified in blood cultures from patients diagnosed with catheter-related bloodstream infections (CRBSIs) during the period.
Methods And Results:
From May 2019 to December 2023, monthly samples were collected from eight designated points in the water treatment system.
Results:
Positive cultures accounted for 10.14% of samples, with 52.87% in the pre-stage and 49.1% in the post-filt stage. Predominant microorganisms included Sphingopyxis spp., Pseudomonas spp., and Brevundimonas aurantiaca. Microbiological growths were not observed in sampling point after ultraviolet radiation (UV). One CRBSI due to Pseudomonas was reported, with no temporal link between isolation in blood and water.
Conclusions:
Despite low pathogenicity in most cases, microbiological control in water needs strict monitoring and effective water treatment systems. Enhanced procedures such as UV light filtration should be considered to minimize risks in patients.
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Hemodialysis III: Nursing Management
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Microbial Wastewater Treatment