Performance evaluation of low-volume blood collection tubes for coagulation tests
Qingna Zhai1, Shixiao Chen1, Haijun Xiao1
1Department of Laboratory Medicine, Shenzhen City Baoan District Women's and Children's Hospital, 56 Yulv Road, Baoan District, Shenzhen, Guangdong, 518102, China.
Background:
Frequent phlebotomy for laboratory testing is a major cause of iatrogenic blood loss, particularly in critical care settings. Conventional large-volume blood collection tubes often collect blood far in excess of analyzer requirements. This study evaluated the analytical performance of novel 1 mL vacuum blood collection tubes compared to conventional 3 mL tubes for coagulation testing, assessing their potential to reduce diagnostic blood loss without compromising test quality.
Methods:
Following the established guidelines (CLSI EP07 and WST 224-2018), the performance of 1 mL microsample tubes was systematically compared with conventional 3 mL tubes. The evaluation indicators covered aspiration volume accuracy, result comparability of six coagulation parameters, and anti-interference capability against hemolysis, icterus, and lipemia (HIL). The six detected coagulation parameters included activated partial thromboplastin time (APTT), prothrombin time (PT), thrombin time (TT), fibrinogen (FIB), D-dimer, and antithrombin III (ATIII). Passing-Bablok regression analysis and Bland-Altman analysis were performed to assess the consistency and bias of detection results. All tests were implemented on a CX9000 automated coagulation analyzer.
Results:
The aspiration volume deviation (Rb) for the microsample tube (4.0%) was comparable to the conventional tube (4.9%), both within the acceptable ± 10% tolerance. Passing-Bablok regression demonstrated high concordance between 1 mL microtubes and conventional tubes across all six coagulation parameters (n = 46-71), with slopes ranging from 0.931 to 1.051 and intercepts from - 0.018 to 2.555; all P-values for deviation from linearity were non-significant. Bland-Altman analysis confirmed that all evaluated coagulation parameters met the predefined clinical consistency criteria.
Conclusion:
The 1 mL vacuum coagulation tube demonstrated equivalent analytical performance to conventional 3 mL tubes across 6 routine coagulation parameters, including aspiration accuracy, result comparability, and interference resistance. Its implementation can significantly reduce blood draw volumes by over 50%, offering a reliable and blood-conserving solution for coagulation testing, especially beneficial for critical care, pediatric, and other vulnerable patients.