Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Feasibility of Using Machine Performance Check-enhanced Couch Check as a Daily Surrogate for Multisphere Winston-Lutz
Tae Kyu Lee1,2
1Department of Radiation Oncology, Allina Health Cancer Institute, Minneapolis, MN, USA.
Journal of Medical Physics
|July 9, 2026
Summary
Varian
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiosurgery Technology
Background:
- Single-isocenter multitarget stereotactic radiosurgery (SIMT SRS) demands high geometric accuracy.
- Winston-Lutz (WL) testing is the established standard for stereotactic verification but is time-consuming.
- Alternative methods for routine geometric accuracy checks are needed to streamline clinical workflows.
Purpose of the Study:
- To assess Varian's machine performance check (MPC)-enhanced couch check as a potential substitute for WL testing.
- To compare the geometric accuracy measurements between WL tests and MPC-enhanced couch checks in clinical practice.
- To determine if MPC can reliably verify isocenter and couch rotation within SIMT SRS tolerances.
Main Methods:
- Geometric agreement was evaluated between WL tests and MPC-enhanced couch check data.
- WL tests were conducted using a multisphere VisionRT phantom over 23 SRS/SRT treatment days.
- Pearson correlation and Bland-Altman analysis were used to compare WL measurements with MPC-max and MPC-shift values.
Main Results:
- High correlation (r > 0.85) was observed between WL and MPC metrics.
- MPC-max and MPC-shift showed minimal bias (+0.033 mm and +0.037 mm, respectively).
- The 95% limits of agreement were within ±0.25 mm, meeting SIMT SRS geometric tolerances.
Conclusions:
- MPC-enhanced couch check demonstrates strong agreement with WL measurements.
- MPC serves as a practical and accurate surrogate for WL testing in SIMT SRS.
- This supports the integration of MPC for frequent isocenter and couch-rotation verification in clinical SRS workflows.