Contact allergy to a titanium port catheter: Diagnostic challenges and clinical resolution
Silas Fugger1, Alexander Kauffmann1,2, Heinrich Dickel1
1Department of Dermatology, Venereology and Allergology, St. Josef Hospital, University Medical Center, Bochum, and.
Background:
Titanium and its alloys are widely used in modern medical devices because of their favorable biocompatibility and mechanical properties. Allergic reactions to titanium-based implants are considered rare, and their diagnosis is hampered by the poor dermal penetration of metallic allergens.
Case Report:
A 29-year-old female patient with seronegative myasthenia gravis requiring parenteral nutrition presented with persistent inflammatory dermatitis overlying a titanium port catheter (X-Port BARD Titan low-profile) 4 weeks after implantation in July 2023. Despite initial diagnostic uncertainty and sequential management modifications, her condition continued to deteriorate. Similar eruptions developed after re-implantation of the same titanium port model on the contralateral thorax. Rigorous patch testing incorporating standardized tape-stripping and extended reading intervals revealed positive sensitization to the titanium port body, while all other port components and an alternative plastic port system (BARD SlimPort M.R.I. Ultra Low-Profile) tested negative. Energy-dispersive X-ray spectroscopy confirmed the titanium body composition as a Ti-Al6-V4 alloy. Substitution with the plastic port catheter in February 2025 resulted in complete symptom resolution, with sustained clinical remission through July 2025.
Conclusion:
Although very uncommon, contact sensitization to titanium medical devices can occur and may present significant diagnostic and therapeutic challenges. Rigorous patch testing methodology, incorporating standardized tape-stripping and extended 168-hour readings, appears essential for detecting delayed-type hypersensitivity to metallic implants. Clinical resolution following biocompatible device substitution supports an allergic etiology in this case.
Related Concept Videos
Endocarditis IV: Nursing Management
Drug Toxicity: Allergic Reactions
Cardiac Catheterization II: Right Heart Catheterization
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Cardiac Catheterization I: Pre-Procedure Overview
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...