Accuracy of Toric Intraocular Lens Realignment Calculators: A Vector-Based Bivariate Comparison of Current
Klemens P Kaiser1, Jascha A Wendelstein2, Mariam Sobhi1
1From the Department of Ophthalmology (K.P.K., M.S., P.N., T.K., C.M.L.), Goethe-University, Frankfurt am Main, Germany.
American Journal of Ophthalmology
|July 11, 2026
Summary
Three toric intraocular lens (IOL) calculators (AstiFix, Gatinel-TOR, Ophtera-AA) offer consistent axis recommendations for toric IOL realignment, showing no systematic bias. Other methods like TorAlign and IOLCon-LPC had divergent recommendations.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optics
Background:
- Toric intraocular lenses (IOLs) correct astigmatism but require precise axis alignment.
- Residual astigmatism after cataract surgery necessitates IOL realignment.
- Accurate prediction of astigmatic error is crucial for successful IOL outcomes.
Purpose of the Study:
- To compare the accuracy of available calculation methods for toric IOL realignment.
- To evaluate astigmatic prediction error (PE) using vector and bivariate analysis.
- To assess the consistency of recommended rotation axes for toric IOLs.
Main Methods:
- Retrospective comparative accuracy study involving 49 eyes.
- Calculated predicted post-realignment astigmatism and rotation axes using AstigmatismFix, Gatinel-TOR, Ophtera-AA, TorAlign, and IOLCon-LPC.
- Evaluated astigmatic PE via vector magnitude and bivariate analysis; assessed axis agreement using intraclass correlation coefficients (ICC).
Main Results:
- AstiFix, Gatinel-TOR, and Ophtera-AA demonstrated highly consistent axis recommendations (ICC ≥ 0.997) and low mean vector PE (0.43-0.51D).
- TorAlign showed significantly different axis recommendations (ICC=0.816-0.822), and IOLCon-LPC had the lowest agreement with realignment-specific tools.
- No tested formula showed statistically significant deviation from the zero-vector; repeated pseudophakic measurements did not improve accuracy.
Conclusions:
- Realignment-specific calculators (AstiFix, Gatinel-TOR, Ophtera-AA) using pre-realignment subjective refraction provide reliable and consistent axis recommendations.
- IOLCon-LPC and TorAlign exhibited divergent axis recommendations and higher total variance, indicating less reliability for toric IOL realignment planning.
Related Concept Videos
Accuracy, limits, and approximation
Accuracy, limits, and approximations are common in many fields, especially in engineering calculations. These concepts are imperative for ensuring that a given value is as close as possible to its true value.
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Net Torque Calculations
When a mechanic tries to remove a hex nut with a wrench, it is easier if the force is applied at the farthest end of the wrench handle. The lever arm is the distance from the pivot point (the hex nut in this case) to the person’s hand. If this distance is large, the torque is higher. Only the component of the force perpendicular to the lever arm contributes to the torque. Therefore, pushing the wrench perpendicular to the lever arm is more advantageous. If multiple people apply force to rotate...