Related Experiment Video
Updated: Jul 13, 2026

Using an Automated Hirschberg Test App to Evaluate Ocular Alignment
Published on: March 24, 2020
Development and Clinical Validation of a Digital Eye-Tracking-Based Cover Test for Ocular Misalignment
Eduardo Esteban-Ibañez1, David Solanas1, Marta Lacort-Beltrán1
1Aragon Institute for Health Research (IIS Aragón), Zaragoza, Spain.
The DIVE Cover Test (DCT), a portable digital eye-tracking tool, accurately detects ocular misalignment, showing strong agreement with the classic cover test (CCT). This innovative device offers excellent repeatability and is a practical option for clinical use.
Area of Science:
- Ophthalmology and Vision Science
- Medical Technology and Instrumentation
- Diagnostic Accuracy Studies
Background:
- The cover test is a fundamental clinical assessment for detecting ocular misalignment (strabismus).
- Traditional cover tests rely on subjective observation, introducing potential variability.
- There is a need for objective, portable, and accurate tools to quantify ocular deviations.
Purpose of the Study:
- To develop and validate the DIVE Cover Test (DCT), a portable, eye-tracking-based digital cover test.
- To assess the DCT's diagnostic accuracy, agreement with the classic cover test (CCT), and repeatability.
- To evaluate performance across different cover test phases (cover-uncover test [CUT] and alternate cover test [ACT]) and ocular axes (horizontal and vertical).
Main Methods:
- A prospective clinical validation study involving 68 participants (3-74 years) with and without strabismus.
- Participants underwent repeated DCT and CCT examinations by experienced clinicians at near distance.
- Diagnostic performance (sensitivity/specificity), intermethod agreement (Bland-Altman, ICCs, Pearson correlation), and test-retest repeatability were analyzed.
Main Results:
- The DCT demonstrated high diagnostic accuracy: 85% sensitivity and 98% specificity in CUT; 92% sensitivity and 87% specificity in ACT.
- Strong agreement was observed between DCT and CCT, particularly for horizontal deviations during ACT (ICC=0.96).
- Excellent DCT test-retest repeatability (ICC ≥0.93) and narrow limits of agreement (±3.8 to ±4.6 PD) were achieved across all conditions.
Conclusions:
- The DIVE Cover Test (DCT) is a highly accurate and repeatable tool for quantifying ocular misalignment.
- It shows strong agreement with the established classic cover test (CCT), supporting its clinical utility.
- The DCT offers a practical, portable, and objective alternative for strabismus evaluation in clinical settings.