Hounsfield unit threshold to predict lumbar fusion failure: a finite element study
Dmitrii Sidorenko1, Aleksandr Dol1, Leonid Bessonov1
1Laboratory of digital medical technologies, Saratov State University, Saratov, Russian Federation.
Summary
A Hounsfield unit (HU) value of approximately 125 indicates a critical threshold for lumbar fusion stability. Below this value, trabecular bone strength may be compromised under various loading conditions, increasing the risk of fusion failure.
Area of Science:
- Biomechanical Engineering
- Orthopedic Surgery
- Medical Imaging
Background:
- Clinical studies suggest a link between Hounsfield unit (HU) values and lumbar implant instability.
- Existing HU thresholds lack biomechanical validation for predicting fusion failure.
- A biomechanical rationale is needed to establish reliable HU thresholds.
Purpose of the Study:
- To provide biomechanical evidence for using a Hounsfield unit (HU) threshold to predict lumbar fusion failure.
- To investigate the relationship between vertebral bone density (HU values) and mechanical stress under physiological loading.
- To identify a critical HU value indicative of potential fusion complications.
Main Methods:
- Development of a finite element model of a fused L4-L5 lumbar segment with screws and an interbody cage.
- Numerical simulation of the fused model under standard static loading conditions (flexion, extension, lateral bending, axial rotation).
- Analysis of von Mises stresses in vertebral trabecular bone with varying HU values (90-256).
Main Results:
- Maximum stresses in trabecular bone occurred near screw and cage contact points during extension, lateral bending, axial rotation, and compression.
- At HU 165+, maximum trabecular bone stress remained below ultimate strength.
- At HU 140, stress exceeded ultimate strength during flexion; at HU 125, it exceeded ultimate strength under all four loading conditions.
- Cortical bone and screws did not reach critical damage levels even at HU 90.
Conclusions:
- An HU value of approximately 125 appears to be a transition point for lumbar fusion.
- At HU 125, trabecular bone stresses approach or exceed predicted strength under multiple loading modes.
- Cortical bone integrity was maintained across the tested HU range, suggesting trabecular bone density is the limiting factor.