Related Experiment Video
Updated: Jul 15, 2026

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Design and Technical Validation of a Novel O2 Mixed Fitting for Dual Connection in Medical Oxygen Delivery Systems
Oscar Ivan Quintero Osorio1, Berkan Basançelebi2, Elinaldo da Conceição Dos Santos3
1Centro de Investigaciones Clínicas, Fundación Valle del Lili, Cali, Colombia.
Background:
Medical oxygen delivery systems require adaptable connection interfaces capable of supporting different clinical configurations during oxygen therapy. Integrating multifunctional connection mechanisms within a single component offers an opportunity to improve system versatility and connection management in respiratory care settings.
Objective:
To design and experimentally validate a novel dual-connection fitting for medical oxygen delivery, designated as the O2 Mixed Fitting, that enables simultaneous attachment of both a bubble humidifier and conventional oxygen tubing while remaining installed on the flowmeter.
Methods:
A comprehensive state-of-the-art review informed the conceptual development of the device. Iterative design refinement was performed using computer-aided design. The final device was manufactured by injection molding using medical-grade Acrylonitrile Butadiene Styrene. Validation included gas flow analysis, leak and pressure-holding tests, microbiological and physical stability evaluation under controlled climatic conditions, and sustainability assessment using a structured eco-design framework.
Results:
The integrated dual-connection mechanism enabled transition between humidified and non-humidified oxygen delivery without detachment from the flowmeter. Gas flow measurements remained within predefined accuracy tolerances, including ±0.5 L/min at low flows and ±10% at higher flows. No leakage was detected during pressure testing up to 10 psi. During the 12-month natural stability study, physical and microbiological parameters remained within predefined acceptance criteria under climatic zone IVb conditions. The eco-design assessment demonstrated favorable environmental performance.
Conclusion:
The O2 Mixed Fitting represents an innovative dual-connection solution for medical oxygen delivery applications. Laboratory validation confirmed structural integrity, flow stability, pressure resistance, and physical and microbiological stability under the evaluated conditions. The integrated design enhances functional versatility while maintaining full compatibility with existing clinical infrastructure. Further clinical and operational studies are warranted to evaluate its potential impact on patient care and clinical workflow.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.