Validation of a New Handheld Automated Fixed-Force Applanation Tonometer
Fiona S McDonnell1, Erin Simons2, Michael Zhu3
1Department of Ophthalmology and Visual Sciences, University of Utah, Salt Lake City, UT, USA.
A new handheld tonometer accurately measures intraocular pressure (IOP) in human eyes. This automated device shows high precision, particularly within the clinically relevant 10-30 mm Hg range, suggesting potential for patient care.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Devices
Background:
- Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- Existing tonometry methods have limitations in accuracy and ease of use.
- A novel automated fixed-force applanation tonometer was developed to address these challenges.
Purpose of the Study:
- To validate a novel handheld automated fixed-force applanation tonometer.
- To assess the accuracy of the tonometer using ex vivo human eyes.
- To determine the tonometer's performance across a range of intraocular pressures.
Main Methods:
- A custom tonometer applying constant force was used with a camera to record applanation images.
- Software automatically calculated intraocular pressure (IOP) from the applanation area.
- Ex vivo human eyes were set to IOPs from 5 to 45 mm Hg manometrically, and measurements were compared to set values.
Main Results:
- The novel tonometer demonstrated high accuracy when compared to manometrically set IOPs in ex vivo human eyes.
- Mean absolute difference between measured and set IOP was 3.1 ± 3.5 mm Hg across the full range.
- Accuracy improved significantly in the clinically relevant 10-30 mm Hg range (mean difference 0.7 ± 1.7 mm Hg).
- A strong linear correlation was observed between prototype and set IOP (R² = 0.90, P < 0.0001), improving to R² = 0.95 (P < 0.0001) for 10-30 mm Hg.
Conclusions:
- The novel handheld automated fixed-force applanation tonometer exhibits high accuracy in IOP measurement.
- The device shows particular precision within the 10- to 30-mm Hg IOP range.
- This tonometer has potential applicability for improving patient care in ophthalmology.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure