Related Experiment Video
Updated: Aug 6, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Interventional cardiac magnetic resonance-guided flutter ablation: Technical evolution and procedural insights from
Luuk H G A Hopman1, Sarah Mahmoodi1, Raschel D van Luijk2
1Department of Cardiology, Amsterdam UMC, Amsterdam, The Netherlands.
Background:
Interventional cardiac magnetic resonance imaging (iCMR) enables catheter ablation without ionizing radiation and provides real-time soft-tissue visualization. Recent developments have enabled initial clinical applications, with typical atrial flutter ablation serving as a benchmark to evaluate feasibility.
Objective:
Describe procedural feasibility, efficiency, and clinical outcomes of iCMR-guided cavotricuspid isthmus (CTI) ablation in a single-center cohort.
Methods:
We retrospectively analyzed 35 consecutive patients who underwent real-time iCMR-guided CTI ablation between September 2022 and February 2025. Procedures were performed on a 1.5T MR imaging scanner using active catheter imaging (ACI, n = 15) or active catheter tracking (ACT, n = 20). Procedural characteristics, technical performance, acute success, complications, and 1-year arrhythmia recurrence were evaluated.
Results:
Mean age was 66 ± 7 years, and 9% were women. Acute bidirectional CTI block could be confirmed in 32 of 35 patients (91%). Median (interquartile range [IQR]) procedure duration was 106 (IQR, 70-128) minutes with ACI and 133 (IQR, 94-161) minutes with ACT (P = .07). Procedural or technical challenges occurred in 19 patients (54%), most commonly related to coronary sinus catheter positioning and workflow or system-integration issues. No major complications occurred. During 1-year follow-up, typical atrial flutter recurred in 6 of 32 patients (19%) with confirmed CTI block, with no difference between ACI (4 of 15 patients) and ACT (2 of 17 patients) (log-rank P = .29).
Conclusion:
iCMR-guided CTI ablation was feasible and could be performed safely using both ACI and ACT. However, procedural complexity, prolonged procedure times, and recurrence rates reflect the early implementation phase of this first-generation iCMR platform. Ongoing advances in catheter technology, system integration, and workflow optimization are expected to further improve procedural performance and support broader clinical adoption.