Single-Phase Surrogate for True Non-Contrast Liver CT Using 190-keV Monoenergetic Imaging on Photon-Counting CT:
Takashi Ohtani1, Masato Shimada1, Kaede Nishiyama1
1Radiological Center, University of Fukui Hospital, Japan (T.O., M.S., K.N., T.T.).
Rationale And Objectives:
To evaluate whether 190-keV virtual monoenergetic imaging (VMI) using photon-counting detector CT (PCD-CT) can serve as a surrogate for true non-contrast (TNC) in hepatic steatosis quantification and to compare its attenuation agreement with virtual non-contrast (VNC).
Methods:
Forty-nine patients who underwent abdominal PCD-CT and MRI-proton density fat fraction (PDFF) within 2 months were retrospectively analyzed. VNC and 190-keV VMI were reconstructed from equilibrium-phase scans. Liver attenuation was measured on TNC, VNC, and 190-keV VMI in four hepatic segments. Pearson correlation and Bland-Altman analyses assessed agreement with TNC. Receiver operating characteristic analysis evaluated diagnostic performance for steatosis (MRI-PDFF ≥ 5%). Statistical analyses were performed using MedCalc (version 23.4) and GraphPad Prism (version 10.6.1). Statistical significance was defined as p < 0.05.
Results:
The study included 49 patients (30 men, 19 women; mean age, 66.7 ± 11.2 years). In whole-liver analysis, 190-keV VMI correlated strongly with TNC (r=0.921, p<0.001), comparable to VNC (r=0.925). In the anterior segment, VNC showed a larger positive bias relative to TNC (+5.5 HU), whereas 190-keV VMI showed a smaller bias (+2.2 HU) that more accurately reproduced the regional attenuation pattern. For detecting steatosis in the anterior segment, a significant difference in area under the curve was observed between TNC and VNC (0.921 vs 0.817, p=0.018), whereas no significant difference was found between TNC and 190-keV VMI (0.921 vs 0.869, p=0.149).
Conclusion:
190-keV VMI may serve as a practical surrogate for TNC in hepatic steatosis quantification, demonstrating reduced segment-dependent bias compared with VNC in artifact-prone regions.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
07:47Non-Invasive PET/MR Imaging in an Orthotopic Mouse Model of Hepatocellular Carcinoma
Published on: August 31, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...