The impact of protective eyewear and device position on operator radiation dose during handheld portable X‑ray device
Sinem Coşkun Albayrak1, Alican Kuran2, Murat Doğuş Günel3
1Faculty of Dentistry, Department of Oral and Maxillofacial Radiology, Lokman Hekim University, Ankara, 06510, Turkey.
Objectives:
The use of handheld dental X-ray devices has been increasing, raising concerns about operator radiation exposure. This study evaluated the effects of handheld dental X-ray device positioning and protective eyewear on the absorbed dose received by the operator's eye lens and pituitary gland.
Methods:
Calibrated optically stimulated luminescent dosimeters were placed at the eye lens and sphenoid regions of an anthropomorphic head phantom. Measurements were obtained under different device positioning scenarios, with and without protective eyewear. Four operator positions were evaluated, and 30 repeated irradiations were performed under each condition, both with and without protective eyewear, resulting in a total of 720 individual measurements.
Results:
Both operator positioning and the use of protective eyewear significantly influenced absorbed dose levels. Protective eyewear reduced ocular and sphenoidal exposure across all evaluated positions, with the greatest reduction observed when combined with optimal device angulation. Across all positions, the use of protective eyewear decreased absorbed dose by approximately 48-70% (p < 0.001 for all comparisons), with the largest reductions observed under the shielding condition.
Conclusions:
Appropriate operator positioning and the consistent use of protective eyewear are essential for minimizing radiation exposure to radiosensitive structures during handheld dental X-ray procedures. Under the conditions evaluated in this phantom study, the combination of a 45° device angulation and protective eyewear resulted in the lowest absorbed dose values, suggesting that this approach may help reduce operator radiation exposure.
Clinical Trial Number:
Not applicable.
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
PPE Use in Healthcare Settings I: Donning
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...