Related Experiment Video
Updated: Aug 6, 2026

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
Flexible metasurface boosts performance of bio-inspired surface coil for 7 T preclinical MRI
S E Solis-Najera1, J Lazovic2, A Villarreal1
1Departamento de Fisica, Facultad de Ciencias, UNAM, Mexico City 04510, Mexico.
Abstract:
This work presents a bio-inspired, floral-patterned surface coil for ultra-high-field preclinical MRI, enhanced by a flexible split-ring metasurface that boosts SNR without passive tuning. When the coil is loaded with a saline-solution phantom, RF impedance matching is degraded, leading to a reduction in performance. To quantify this degradation, we evaluated the noise factor (NF) and SNR from phantom images. The NF ratios demonstrate the metasurface's advantages: without metasurface, NFcir/NFbio=1.02 (where NFcir and NFbio correspond to the circular and bio-inspired coils, respectively); with metasurface and phantom, NFcirMS/NFbioMS=3.52 (where the subscript MS denotes the addition of the metasurface), indicating drastic noise reduction. SNR results show that the bio-inspired coil alone provides a 28% improvement over a conventional circular coil, while adding the metasurface yields an additional 44% enhancement (84% overall). These results highlight the potential of combining bio-inspired geometries with flexible metasurfaces for high-sensitivity ultra-high-field MRI.