Epicardial Lead Dysfunction in a Patient With Mechanical Tricuspid Valve-A Challenging Complication
Tuba Guctekin1, Mucahit Yasin Guctekin2
1Department of Cardiology, Baltalimanı Bone Diseases Training and Research Hospital, Istanbul, Turkey.
None:
Complete atrioventricular block after mechanical tricuspid valve replacement represents a challenging pacing scenario because transvalvular right ventricular lead placement is contraindicated. We report a 51-year-old woman with mechanical mitral and tricuspid valves who presented with syncope and complete AV block due to late epicardial lead dysfunction. Transient restoration of capture during transthoracic echocardiography suggested dynamic lead-myocardial contact or intermittent lead continuity. Coronary sinus pacing was attempted but abandoned because of intolerable diaphragmatic stimulation. The patient later deteriorated and died from septic shock. This case highlights the limitations of available pacing strategies after mechanical tricuspid valve replacement.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse I: Introduction
Endocarditis III: Medical Management
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiomyopathy II: Dilated Cardiomyopathy
Aortic Regurgitation I: Introduction