Related Experiment Video
Updated: Aug 5, 2026

Controlling Parkinson's Disease With Adaptive Deep Brain Stimulation
Published on: July 16, 2014
Charging failure due to pulse generator rotation after deep brain stimulation: a case report
Di Lu1,2, Mingming Zhao1,2, Feng Yin1,2
1Department of Neurosurgery, Aerospace Center Hospital, No. 15 Yuquan Road, Haidian District, Beijing 100049, China.
Abstract:
A 54-year-old man with Parkinson's disease presented with gait instability and difficulty charging his rechargeable deep brain stimulation (DBS) pulse generator. Chest radiography revealed horizontal rotation and anterior-to-posterior inversion of the device within the subcutaneous pocket, which was not seen in prior imaging. Surgical revision confirmed the device malorientation, and repositioning restored charging efficiency. This case highlights that pulse generator rotation, a rare but relevant hardware complication, can cause charging difficulties despite normal device interrogation. It underscores the importance of considering mechanical causes and using chest imaging in postoperative assessments.