Comparative analysis of 6-lead and single-lead consumer-grade electrocardiograms: diagnostic accuracy, numerical
Myoung Jung Kim1, Sujeong Eom2, Juwon Kim3
1Division of Cardiology, Department of Internal Medicine, Seoul St Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
Aims:
Consumer-grade electrocardiogram (ECG) devices enable accessible rhythm monitoring. Recently developed six-lead handheld ECGs promise improved signal quality and diagnostic performance; however, comparative validation remains limited. Using 12-lead ECGs as reference, we compared diagnostic performance and quantitative agreement of a 6-lead and single-lead consumer ECGs.
Methods And Results:
In this prospective single-centre study, 194 arrhythmia-clinic patients (498 paired recordings) underwent simultaneous 10-s 12-lead and 30-s 6-lead ECG acquisition, followed by sequential 30-s single-lead recording. Two blinded electrophysiologists interpreted ECGs. Diagnostic performance for predefined rhythm categories (sinus rhythm, atrial/ventricular premature complex, atrial fibrillation/flutter/tachycardia, atrioventricular block, others) was assessed by sensitivity, specificity, and accuracy. Waveform agreement was evaluated using Bland-Altman analysis and intraclass correlation coefficients (ICC) accounting for repeated measures. Sensitivity was higher for the 6-lead ECG for ectopic beats, atrial flutter, and first-degree atrioventricular block, with comparable specificity across categories. Overall diagnostic accuracy was 98.6% (95% confidence interval, 97.4-99.6) for the 6-lead and 96.9% (94.7-98.7) for the single-lead ECG. Agreement with the 12-lead ECG was higher for the 6-lead device for PR interval (ICC 0.89; 0.85-0.92) and QRS amplitude (0.96; 0.94-0.97) than for the single-lead device (0.57; 0.48-0.63 and 0.04; -0.13-0.21, respectively). Bland-Altman analysis demonstrated generally narrower limits of agreement for the 6-lead ECG. Findings were consistent in patient-level sensitivity analyses.
Conclusion:
The 6-lead handheld ECG demonstrated a higher point estimate for diagnostic accuracy and closer agreement with the 12-lead ECG than the single-lead smartwatch ECG, supporting its use for arrhythmia assessment and interval measurement.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...