Ultrasonography-Guided Nasogastric Tube Placement and Verification: A Standardized Nursing Protocol
Published on: April 7, 2026
Effect of a Novel Gastric Tube Insertion Method Using a Light-Emitting Diode Source and Fiber in Neonates: A
Kaoru Okazaki1, Jumpei Kuroda1, Eiji Hirakawa2
1Tokyo Metropolitan Children's Medical Center, Department of Pediatrics, Tokyo, Japan, Fuchu.
American Journal of Perinatology
|July 28, 2026
Summary
A new light-emitting diode source and fiber (LED-SF) method accurately confirms nasogastric tube (NGT) or orogastric tube (OGT) placement in neonates. This safe and effective technique offers reliable bedside verification for preterm and term infants.
Area of Science:
- Neonatal Medicine
- Medical Devices
- Clinical Procedures
Background:
- Blind insertion of nasogastric tubes (NGT) or orogastric tubes (OGT) can lead to inaccurate stomach placement.
- Accurate verification of NGT/OGT tip position is crucial for neonate safety and effective treatment.
Purpose of the Study:
- To evaluate the accuracy and safety of a light-emitting diode source and fiber (LED-SF) system for confirming NGT/OGT tip placement in neonates.
- To assess the reliability of LED-SF in determining correct tube positioning compared to standard methods.
Main Methods:
- Prospective multicenter study involving 101 neonates requiring NGT/OGT insertion across five tertiary NICUs in Japan.
- Utilized LED-SF technology for initial tip placement verification and compared measured length to post-X-ray determination.
- Statistical analysis included cross-tabulation tests and linear regression.
Main Results:
- 100% of NGT/OGT tips were confirmed to be in the stomach using the LED-SF method.
- No adverse events were reported during the study.
- A significant correlation (r2 = 0.844, P<0.001) was found between insertion length measured by LED-SF and final length confirmed by X-ray.
Conclusions:
- The LED-SF method is an effective, safe, and easily usable tool for verifying NGT/OGT tip placement in neonates.
- LED-SF shows potential as a reliable bedside standard for confirming tube placement in both preterm and term infants.
- Minimal training is required for healthcare providers to utilize this technology effectively.