Influence of Sample Preparation Methods for Porcine Tissue: Calibration-Free LIBS for In Vivo Applications
Edith Böhmer1,2, Oana-Maria Thoma3,4, Maximilian J Waldner3,4
1Institute of Photonic Technologies (LPT), Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Journal of Biophotonics
|July 29, 2026
Summary
Calibration-free laser-induced breakdown spectroscopy (CF-LIBS) is effective for analyzing untreated biological tissues. Immediate analysis of fresh samples is crucial to avoid elemental shifts from post-mortem changes.
Area of Science:
- Analytical Chemistry
- Biomedical Spectroscopy
Background:
- Calibration-free laser-induced breakdown spectroscopy (CF-LIBS) offers simplified sample analysis.
- Biological tissue sample preparation can influence spectral data in LIBS analysis.
Purpose of the Study:
- To evaluate the impact of different preparation methods on porcine tissue analysis using CF-LIBS.
- To determine the reliability of CF-LIBS for analyzing untreated biological samples.
Main Methods:
- Porcine tissue samples were prepared by embedding in PBS or deionized water, cryo-sectioning, or left untreated.
- Samples were analyzed using calibration-free laser-induced breakdown spectroscopy (CF-LIBS).
- Elemental intensity ratios (e.g., Mg/Na) were compared across preparation methods.
Main Results:
- Embedding porcine tissue in aqueous solutions (PBS, deionized water) altered LIBS spectra, particularly minor deviations in Mg/Na ratios in skin.
- Ionic exchange between embedded samples and immersion media can affect elemental signatures.
- Untreated samples showed spectral shifts due to post-mortem changes, highlighting the need for immediate analysis.
Conclusions:
- CF-LIBS can be reliably applied to untreated biological samples, minimizing sample preparation.
- Immediate analysis of fresh biological tissues is essential for accurate elemental quantification.
- CF-LIBS shows promise for in vivo diagnostics due to its minimal sample preparation requirements.