Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems
Fabian Andres Castaño1,2, Alher Mauricio Hernández3
1Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia. fabian.castano@udea.edu.co.
Background And Objective:
Chronic cardiovascular diseases have motivated the development of wearable systems capable of continuously monitoring physiological variables such as oxygen saturation (SpO ) and non-invasive blood pressure (NIBP). However, motion artifacts remain one of the main limitations affecting the reliability of these measurements in ambulatory environments. This article presents a novel framework for motion artifact reduction in photoplethysmography (PPG) and NIBP signals based on redundant sensing, inertial measurements, ECG-guided temporal synchrony, and multichannel signal processing.
Methods:
The proposed framework combines independent component analysis and recursive least-squares adaptive filtering to separate physiological information from motion-induced interference. Performance was compared with conventional finite impulse response (FIR) filtering and wavelet shrinkage (WS) methods using signal-to-noise ratio (SNR), weighted distortion assessment (WDA), oxygen saturation estimation, and blood pressure estimation metrics.
Results:
The proposed method achieved statistically significant improvements in signal quality and physiological parameter estimation. Average SNR improved from dB to dB in the red PPG channel ( ), while the WDA index increased from 0.48 to 0.73. Oxygen saturation estimation improved from in motion-contaminated signals to after denoising ( ), approaching resting measurements ( ). For NIBP signals, statistically significant improvements were also observed ( ), particularly for systolic pressure estimation, although diastolic pressure remained affected by residual motion-related distortions.
Conclusions:
The proposed framework improves the robustness of wearable PPG and NIBP monitoring under motion conditions through the combined use of sensor redundancy, ECG-guided temporal coupling, inertial measurements, and adaptive multichannel signal processing. The results demonstrate significant improvements in signal quality and physiological parameter estimation compared with conventional denoising approaches, supporting the potential use of the method in ambulatory cardiovascular monitoring applications.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...