Image-based evaluation of a commercial AI synthetic CT generator for brain and prostate MR-only radiotherapy
Morgan Aire1, Carson Matthews2, Joshua Jones1
1Department of Physics and Astronomy, Louisiana State University, Baton Rouge, Louisiana, USA.
Background:
Synthetic CTs (sCT) were developed to provide electron density information for MR-only treatment planning. Some AI-based sCT generators have received FDA-clearance and better capture tissue heterogeneity through voxel-wise mappings to improve image guidance and dose calculation.
Purpose:
To quantitatively characterize the image quality and geometric surrogate agreement of an FDA 510(k) cleared AI-based synthetic CT generator (TheraPanacea) for brain and prostate MR-only radiotherapy, using per-structure volumetric analysis, Hounsfield unit (HU) assessment, and multi-metric geometric evaluation.
Methods:
Paired CT/MR datasets from 20 brain and 20 prostate patients were retrospectively analyzed. sCTs were generated from T1w (brain) and T2w (pelvis) MR images acquired on the same day with identical immobilization. Auto-contours from TheraPanacea and MIM Contour ProtégéAI were compared against MIM Contour ProtégéAI contours on the reference CT. Bone was segmented via a fixed-threshold method. Evaluation metrics included contour volumes, cross-contour HU comparisons, centroid offsets, overlap and surface-based similarity metrics, and manual bony landmark measurements.
Results:
Soft-tissue mean HU differences were <50 HU across both cohorts. Central ROIs showed <20 HU deviations, apart from heterogeneous bone and air-filled regions. Bony structures exhibited HU range compression in the sCT, where HU differences between the sCT and the reference CT differed by a maximum of 347 ± 14 HU in the femoral heads and 455 ± 50 HU in the sacrum. Global histograms confirmed HU range compression. Centroids shifted < 2 mm for all contours (brain A-P predominant, pelvis non-systematic). Boundary metrics showed MDA < 5 mm, and bony landmarks differed < 1 mm between the reference CT and the sCT. Small structures in the brain had inconsistent auto-segmentation, displaying different volumes and low Dice/JI scores. Dental implants in the sCT produced localized MR-like susceptibility artifacts. Within manual contoured implant ROIs, the mean HU values were underestimated by ∼9,000 HU relative to CT.
Conclusion:
TheraPanacea sCT demonstrated acceptable soft-tissue HU (± 50 HU) and geometric surrogate agreement (< 2-5 mm) relative to CT, with larger errors in cortical bone, air interfaces, and implants. Site-specific QA must verify bone HU compression, peripheral geometric agreement via landmark measures, and manually review small/high-contrast contours (optic nerves, cochleae, rectum). Institutional image and dosimetric validation, particularly with implanted devices, remains essential prior to clinical deployment.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:54MR Molecular Imaging of Prostate Cancer with a Small Molecular CLT1 Peptide Targeted Contrast Agent
Published on: September 3, 2013
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...