A Comparison of Non-Invasive Methods of Blood Pressure Measurement in Healthy Term Neonates
Ijeoma GeneveveEzeasouba1, Frances SamOkpokowuruk1, Eno EtimNyong1
1Department of Paediatrics, University of Uyo Teaching Hospital, Uyo, Akwa Ibom State.
Background:
Invasive Blood Pressure (BP) monitoring in neonates is fraught with many complications and is rarely practical for routine monitoring, hence the need for alternative methods of BP measurements, especially in well neonates. This study aimed to compare BP measurements in neonates using Oscillometric and Doppler Ultrasound methods.
Methodology:
One hundred and eighty-five neonates were consecutively recruited. Blood pressure measurements were taken at 12, 24, 36, and 48 hours of life using Doppler ultrasound with Aneroid Sphygmomanometer (DAM) (gold standard), Doppler ultrasound with Hybrid Sphygmomanometer (DHM), and oscillometric methods. Data was analysed using the IBM statistical package for the social sciences (SPSS) version 23. Comparison between the mean BP values obtained from different instrumentswas done using the paired Student t-test, and the degree of agreement between values was tested using the intra-class correlation coefficient scores. Bland-Altman analyses were also done. P-value was set at <0.05.
Results:
Blood pressure increased gradually from 12 to 48 hours of life. The blood pressure values using the automated oscillometric method were significantly higher when compared to the values obtained using the gold standard (p < 0.001), and the level of absolute agreement between the two methods was poor (intra-class coefficient < 0.5). Values from the DHM were not significantly different from those from thegold standard method (p > 0.05), and the level of absolute agreement between the two methods was excellent (Intra-class correlation coefficient >0.9).
Conclusion:
There was poor agreement between BP values using oscillometry when compared with DAM. However, the use of DAM or DHM demonstrated no difference in BP values. Caution should be exercised when interpreting values from oscillometric BP measurements in the newborn. On the other hand, either of the Doppler ultrasound methods may be used interchangeably.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure