Assessment of weight-based variations in effective dose for common PET/CT scans in a tertiary hospital
Abdallah T Al Said1, Azhar A Rahman1, Laith K Albadarneh2
1School of Physics, Universiti Sains Malaysia, 11800 USM Minden, Penang, Malaysia.
Background:
Positron Emission Tomography/Computed Tomography (PET/CT) is a crucial imaging modality widely used for diagnosing and staging various cancers. Although patient weight is known to influence radiation dose, there is limited research on optimizing PET/CT protocols based on weight categories to minimize unnecessary exposure.
Objectives:
This study evaluates the effective dose for common PET/CT scans at Penang Hospital, with a particular focus on the influence of patient weight on both PET and CT dose components. Additionally, the study assesses whether weight-based dose protocols can contribute to dose optimization.
Methods:
This retrospective study analyzed 1439 PET/CT scans performed at Penang Hospital. The effective dose was calculated using International Commission on Radiological Protection (ICRP) 103, 106, and 128 dose coefficients. Conversion factors used were 0.019 mSv/MBq for 18F-FDG, 0.0199 mSv/MBq for 68Ga-PSMA, and 0.021 mSv/MBq for 68Ga-DOTATATE. The CT dose component was estimated using the dose-length product (DLP) with an empirical conversion factor (k = 0.015 mSv·mGy-1·cm-1) for whole-body CT.
Results:
The total effective dose for 18F-FDG (head-to-thigh and head-to-feet) scans was 12.34 and 16.38 mSv, respectively. For 68Ga-PSMA and 68Ga-DOTATATE scans, the total effective dose was 11.55 and 12.00 mSv, respectively. A statistically significant increase in dose was observed in heavier patients due to increased administered activity and CT parameters (P < .05). Weight had a smaller impact on PET dose in 68Ga-PSMA and 68Ga-DOTATATE compared to 18F-FDG.
Conclusion:
Patient weight significantly influences the total effective dose in PET/CT examinations. These findings highlight the importance of considering patient-specific factors when evaluating radiation dose in clinical practice.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...