Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
Published on: January 5, 2024
Characterizing electronic noise interference from the external laser positioning system on a radiation therapy MRI
Lucas McCullum1,2, Yao Ding3, Clifton D Fuller2
1UT MD Anderson Cancer Center UTHealth Houston Graduate School of Biomedical Sciences, Houston, Texas, USA.
Magnetic resonance imaging (MRI) simulation lasers can degrade image quality by introducing electronic noise, especially with the integrated body coil. Researchers recommend avoiding laser use during MRI scans to maintain image integrity and accurate quantitative measurements.
Area of Science:
- Medical Imaging
- Radiation Oncology
- Medical Physics
Background:
- Magnetic Resonance Imaging (MRI) simulation is crucial for radiation therapy planning, utilizing external laser positioning systems (ELPS) for patient setup.
- ELPS, if active during MRI scans, can introduce electronic noise, potentially degrading image quality.
Purpose of the Study:
- To investigate the impact of MRI simulation lasers on general image quality.
- To assess the effect of these lasers on quantitative imaging accuracy in various MRI sequences.
Main Methods:
- Cine acquisition visualized real-time signal-to-noise ratio (SNR) changes with ELPS activation/deactivation.
- SNR was measured using the American College of Radiology (ACR) protocol across multiple MRI coils and phantoms.
- Quantitative MRI scans assessed T1, T2, ADC, PDFF, and R2* values with and without ELPS activation.
Main Results:
- Visible electronic noise and a two-fold SNR decrease occurred with the integrated body coil when ELPS was active.
- No significant SNR drop was observed with other tested coils.
- Image intensity uniformity, ghosting, and low contrast detectability degraded with the Head/Neck coil; geometric distortion was unaffected.
- Quantitative MRI values showed concordance, but ADC vials had increased standard deviation with ELPS activation.
Conclusions:
- Activation of ELPS during MRI imaging should be avoided to prevent unnecessary image noise.
- This is critical for quality assurance testing, particularly with the integrated body coil, which is highly susceptible to ELPS-induced noise.
- Clinical guidelines are needed to inform staff about ELPS-induced noise interference during MRI procedures.