Is ultrasonic speed of sound estimation for liver fat quantification affected by measurement depth?
Halit Nahit Şendur1, Mehmet Ali Yeşilçimen2, Mahi Nur Cerit2
1Department of Radiology, Gazi University Faculty of Medicine, Ankara, Turkey. hsendur@gazi.edu.tr.
Purpose:
Ultrasonic speed of sound (SS) estimation is one of the techniques that can be used to assess hepatic fat content. Current literature lacks of information for ultrasonic SS estimation technique in terms of depth dependence. We aimed to evaluate the impact of measurement depth on SS measurements.
Methods:
A total of 90 patients who diagnosed with metabolic dysfunction-associated steatotic liver disease (MASLD) were included in the study. Upper edge of the region of interests (ROI) were positioned at 4 cm and 6 cm deep from the skin. Five measurements were obtained for both depths and the median values were noted for the analysis. Paired-t test was used to compare the measurements.
Results:
The means of the SS measurements for 4 cm and 6 cm depths were 1514 ± 13.5 m/s and 1512.2 ± 15.1 m/s, respectively. There was a significant difference between the SS measurements obtained at 4 cm and 6 cm depths (P = 0.018). Subgroup analyses for BMI (< 30 kg/m2 vs. ≥ 30 kg/m2) and skin to liver capsule (SLC) distance (< 25 mm vs. ≥ 25 mm) revealed that there were significant differences in SS measurements between 4 cm and 6 cm depths in patients with high BMI (P = 0.047) and high SLC (P = 0.044), while there were no significant differences detected in patients with low BMI (P = 0.198) and low SLC (P = 0.140).
Conclusion:
Our results revealed that ultrasonic SS measurements are affected by the measurement depth in patients with high BMI and high SLC. These findings may be valuable for technical standardization of SS technique.
More Related Videos
07:13Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
07:38Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU)
Published on: November 3, 2015
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...