Related Experiment Video
Updated: Aug 12, 2026

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Comparative Performance of Oxygenators for Extracorporeal Membrane Oxygenation in Adults: A Retrospective Analysis
Linda E Sousse1,2, Junya Hagiwara1, Qianqian Liu3
1From the Institute for Extracorporeal Life Support, San Antonio, Texas.
Abstract:
Extracorporeal membrane oxygenation (ECMO) is a vital therapy for patients with severe cardiac and pulmonary failure. Despite technological advances, oxygenator dysfunction remains common, yet comparative clinical data across devices are limited. This study evaluated the performance of four contemporary adult oxygenators used in ECMO in a clinical, real-world setting. This retrospective single-center study included adult patients who received venovenous (VV) or venoarterial (VA) ECMO between January 2020 and December 2023 (N = 282). Oxygenators evaluated included the Medtronic Nautilus/Nautilus Smart, Getinge Heart Lung Support Module Advanced 7.0, Getinge Quadrox-iD Adult, and Fresenius X-Lung. Analyses were performed using propensity weighting and included adjustments for Acute Physiology and Chronic Health Evaluation II severity scores and baseline Charlson Comorbidity Index (CCI). No statistically significant differences in oxygen transfer across the membrane lung (VO2) were observed among oxygenator types after adjustment for covariates. VO2 declined over time during support (p < 0.001) and was higher in VV than VA ECMO mode (p = 0.002). Higher CCI was associated with lower VO2 (p = 0.046). Finally, coagulation and hemolysis markers were comparable across all devices. Current-generation oxygenators provide similar gas exchange efficiency and hemocompatibility in clinical practice. Future prospective and multicenter studies incorporating direct measurements of gas transfer may further clarify device-specific characteristics and their impact on clinical outcomes.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...