Related Experiment Video
Updated: Aug 12, 2026

Stem Cell Transplantation Strategies for the Restoration of Cognitive Dysfunction Caused by Cranial Radiotherapy
Published on: October 18, 2011
Radiation protection strategies for FLASH radiotherapy accelerator installation in animal housing
Giorgio Russo1,2, Santa Gabriella Bonaccorsi1,3, Renata Leanza2
1Institute of Bioimaging and Complex Biological Systems, National Research Council (IBSBC-CNR), Via Terzo Millennio 1 Cefalù, Palermo 90015, Italy.
Abstract:
This study presents a radioprotection assessment for an ultra-high dose rate accelerator installation in one of the animal housing of the Centre for Advanced Preclinical in vivo Research. The main problem concerns the absence of a bunker while limiting exposure in nearby work areas. The geometry of the accelerator, the building layouts and the distribution of the environmental dose equivalent were simulated with FLUKA.CERN 4-4.1, modeling a 9 MeV electron beam. The results confirmed compliance with the limit of 0.250 mSv/year for unrestricted areas during daytime operation. Night-time operating zone restrictions apply when no staff are present, allowing beam on for quality controls. The effective doses to the population, researchers and workers are less than 1 mSv/year, in accordance with Legislative Decree 101/2020; the accelerator operators are classified as exposed. For small animals, precautionary dose limit of 1 mSv/year was adopted. The use of lead shielding and site-specific safety measures ensure radiological compliance.