Robust phase unwrapping for Dixon magnetic resonance imaging multiresolution pixel clustering and local surface
Shuyu Cheng1, Piaoran Li1, Xufeng Zheng2
1School of Biomedical Engineering, Southern Medical University, Guangzhou, China.
Quantitative Imaging in Medicine and Surgery
|August 12, 2026
Summary
A new multiresolution pixel clustering and local surface fitting (MrCLOSE) method significantly improves water-fat separation in two-point Dixon MRI. This robust phase-unwrapping technique minimizes water-fat swaps, enhancing diagnostic accuracy in clinical imaging.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging (MRI)
- Image Processing
Background:
- Water-fat separation is vital for clinical MRI.
- The Dixon technique faces challenges like water-fat swapping due to magnetic field inhomogeneity and low SNR.
- Existing methods struggle with robustness in complex scenarios.
Purpose of the Study:
- To develop a robust phase-unwrapping method for improved water-fat separation in two-point Dixon MRI.
- Introduce the multiresolution pixel clustering and local surface fitting (MrCLOSE) algorithm.
- Enhance the reliability of water-fat separation in clinical MRI.
Main Methods:
- Constructing a multiresolution pyramid from in-phase and out-of-phase complex MRI data.
- Performing hierarchical phase unwrapping using pixel clustering and local surface fitting (CLOSE) within a coarse-to-fine framework.
- Evaluating MrCLOSE against B0-NICEbd, ROMEO, Graph Cut, and original CLOSE using simulations and in vivo datasets.
Main Results:
- MrCLOSE demonstrated robust phase unwrapping across various SNRs, disconnected regions, and phase changes (up to 10π).
- In simulations, MrCLOSE significantly reduced error ratios compared to CLOSE, especially at higher thresholds (0.67% vs. 43.71%).
- In vivo studies showed a substantially lower water-fat swap ratio for MrCLOSE (3.38%) compared to other methods (e.g., ROMEO 96.34%, Graph Cut 67.61%).
Conclusions:
- The integration of multiresolution strategies into CLOSE (MrCLOSE) enhances robustness and accuracy in phase unwrapping.
- MrCLOSE offers a promising solution for phase-related MRI applications, particularly for clinical water-fat separation.
- This method addresses key limitations of current Dixon MRI techniques.