Manikin-based noninferiority simulation study of Airgency™: A novel automatic bag valve mask device for emergency
Reza Widianto Sudjud1, Jenifer Kiem Aviani1
1Department of Anaesthesiology and Intensive Care, Faculty of Medicine, Padjadjaran University/Dr. Hasan Sadikin General Central Hospital, Bandung, West Java, Indonesia.
Objectives:
This study evaluated the performance of Airgency™, an automated bag valve mask (BVM) ventilation device developed by Institut Teknologi Bandung and Padjadjaran University, Indonesia, using a manikin-based simulation model. Its performance was compared with conventional hand-squeezed bagvalvemask ventilation performed by medical professionals, with a focus on reducing operator-dependent variability and fatigue.
Methods:
A manikin-based simulation was conducted using an adult airway trainer connected to a lung simulator (ASL5000) set to mimic apneic respiratory failure with normal lung mechanics. The Ambu bag, connected to 15 L/min oxygen, was secured to the manikin. Airgency™ operated in normal mode (tidal volume [VT] 400 mL, 15 bpm) for 30 min, repeated 50 times. Seventy-eight anesthesiology residents performed manual BVM ventilation under identical conditions after standardized training. Breathing rate (BR), VT, and peak inspiratory pressure (PIP) were recorded every minute, whereas fatigue was rated using the Visual Analog Scale for Fatigue (VAS-F, 1-10).
Results:
Airgency™ demonstrated superior ventilation stability. BR remained consistent (14.77 ± 0.08 vs. 16.54 ± 7.05 bpm; P < 0.0001), with much lower variability (coefficient of variation [CV] 0.47% vs. 46.1%-121.2%). VT delivery was steady (397.8 ± 3.1 vs. 406.9 ± 85.4 mL; P < 0.0001; CV 2.2% vs. 13.9%-25.7%), and PIP was more uniform (29.4 ± 0.9 vs. 23.1 ± 8.2 cmH2O; P < 0.00001; CV 2.5% vs. 28.3%-41.1%). Operator fatigue increased progressively during manual ventilation (VAS-F 1.83-5.09; P < 0.0001).
Conclusion:
Airgency™ maintained consistent, precise ventilation with minimal variability and eliminated operator fatigue, supporting its potential as a reliable alternative to manual BVM ventilation in emergency care.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...