Related Experiment Video
Updated: Aug 17, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Correction of image-guided radiotherapy and intrafraction motion using ExacTrac dynamic for head and neck cancer
Simona Denisia Iftime Holst1,2, Maria Andersen2, Annette Ross Jakobsen1,2
1Department for Medical Physics, Aalborg University Hospital, Denmark.
Abstract:
Daily cone-beam computed tomography (CBCT) based image-guided radiotherapy (IGRT) with six degrees of freedom (6DOF) is standard for head and neck cancer. We retrospectively analyzed 669 sessions from 20 patients, using CBCT guided 6DOF couch corrections followed by ExacTrac Dynamic adjustments. Surface-guided radiotherapy (SGRT) continuously monitored intrafraction motion, triggering verification images and couch shifts when tolerances were exceeded. Pitch and roll deviations were linked to translational misalignments up to 3 mm. SGRT prompted repositioning in 100 sessions (15%), although nearly half of these movements were false positives. This demonstrates SGRT's value as a complementary, non-invasive tool for intrafraction monitoring.