Related Experiment Video
Updated: Aug 19, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report
Adam C Pasquinelly1, Brandon Sharkey1, Alicia M Nahhas1
1University of Toledo College of Medicine and Life Sciences, Toledo, Ohio, USA, utoledo.edu.
Abstract:
Implant fracture is a rare but devastating complication of total hip arthroplasty (THA). We present the case of a patient with a history of prior acetabular fracture who underwent conversion to a dual mobility (DM) THA and subsequently fractured at the neck of the femoral component 7 years after surgery. At revision, both the femoral and acetabular components were noted to be clinically retroverted, suggesting chronic anterior impingement as the cause of the catastrophic failure.