Related Experiment Videos
Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited
Jagdish Chandarana1, Kinjal Butani2
1Vedant Diagnostic Center, Rajkot, Gujarat, India.
Introduction:
Hemoglobin A1c has a stringent total allowable error (TEa), making high Six Sigma values difficult to attain. In addition, data from ambulatory accredited laboratories are sparse.
Methods:
Internal quality control (QC) data for 2 levels were collected over 8 months on a single-lot Bio-Rad D-10 high-performance liquid chromatography analyzer. Long-term imprecision was the coefficient of variation (s/x¯) × 100 (CV) of all internal QC results combined, capturing between-month variation. Bias was derived from external quality assessment (EQAS) and assigned QC targets, applied as absolute values. The Sigma metric was (TEa ‒ absolute bias) / CV at TEa levels of 6%, 7%, and 8%; the Quality Goal Index distinguished imprecision-driven from bias-driven performance.
Results:
Long-term CVs were 3.04% (level 1) and 1.34% (level 2); EQASs were satisfactory throughout. At a TEa of 6% with EQAS-derived bias, Sigma metrics were 1.19 and 2.70 and, with a level-specific QC target bias, 1.46 and 4.22. At the Clinical Laboratory Improvement Amendments 2025 criterion of 8%, the Sigma metrics were 2.11 and 5.71. The level 1 Quality Goal Index (0.52) identified imprecision as predominant.
Discussion:
Despite satisfactory external proficiency, Sigma metrics fell below 3 at both levels against a 6% TEa and below 2 at the clinically pivotal lower concentration, driven by imprecision rather than bias. No control strategy compensates for this; therefore, we proposed a precision-focused corrective plan that targets between-month instability.
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Quality Control