Impact of Radiation-Attenuating Drape on Exposure in a Contemporary Cardiac Catheterization Laboratory: The ATTENUATE
Giorgio A Medranda1, Dimitrios Bliagos2, Brian C Case3
1St Francis Heart Center at Good Samaritan University Hospital, West Islip, New York.
Background:
Despite recent improvements in radiation safety, interventionalists are increasingly exposed to radiation during cardiac catheterization laboratory (CCL) procedures. The RADPAD was designed as a protective scatter-radiation absorbing shield with early studies demonstrating a 20% to 62% reduction in scatter-radiation. The objective of this study was to examine the impact of the protective scatter-radiation absorbing shield in a large contemporary randomized controlled trial across multiple CCL procedures.
Methods:
The investigator-initiated, prospective, randomized, controlled ATTENUATE (rAdpad proTecTion drapE iN redUcing rAdiaTion Exposure) trial randomized CCL procedures 1:1 to use of the RADPAD vs no use of the RADPAD. The primary outcome of interest was the most proximal operator's dose area product (DAP)-normalized operator dose (E).
Results:
A total of 1000 CCL procedures were randomized 1:1 to use of the protective scatter-radiation absorbing shield (n = 500) vs no use (n = 500). The use of the protective scatter-radiation absorbing shield was associated with a 45.1% relative reduction in the DAP-normalized operator dose compared with no use of the protective scatter-radiation absorbing shield (0.64 vs 0.35 μSv/Gy⋅cm2; P < .0001). Coronary cases (n = 720) saw a 21.9% relative reduction in the DAP-normalized E using the RADPAD (0.33 vs 0.43 μSv/Gy⋅cm2; P < .001), and structural cases (n = 210) saw a 54.4% relative reduction in the DAP-normalized E using the RADPAD (0.30 vs 0.66 μSv/Gy⋅cm2; P = .1094).
Conclusions:
In the largest randomized controlled trial to date evaluating the RADPAD protective scatter-radiation absorbing shield, encompassing contemporary coronary and structural CCL procedures, use of the protective scatter-radiation absorbing shield led to a marked decrease in proximal operator radiation exposure when adjusted for the total radiation delivered during each procedure.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization II: Right Heart Catheterization