Published on: August 18, 2017
Quantification of Proinsulin in Human Plasma by Proteolysis-Aided Peptide Immunoaffinity-Liquid Chromatography-Tandem
Sara K Shijo1, Elisha Goonatilleke1, Lili Barahona-Carrillo1
1Department of Laboratory Medicine & Pathology, University of Washington, Seattle, WA, United States.
Background:
Plasma proinsulin concentrations are used to investigate hypoglycemia. They have also been proposed as a marker of β-cell function, particularly as a ratio with C-peptide. Immunoassays remain the primary method for measuring proinsulin despite potential limitations. Mass spectrometry-based assays have been described, but are semiquantitative or rely on nano-flow liquid chromatography. We aimed to develop a liquid chromatography-tandem mass spectrometry assay (LC-MS/MS) at typical clinical laboratory flow rates to quantify proinsulin and its partially processed forms in human plasma and ensure accuracy over time with distributable well-characterized calibrators.
Methods:
Sample preparation consists of protein precipitation, Glu-C digestion, peptide immunoaffinity enrichment, and LC-MS/MS analysis of 2 surrogate peptides: RGFFYTPKTRREAE spans the cleavage site for des-31,32-proinsulin and GSLQKRGIVE spans the site for des-64,65-proinsulin. A purified protein calibration material was characterized (HPLC and amino acid analysis) and used to value-assign a matrix-matched single-point calibration material.
Results:
Within-batch and between-batch imprecision were ≤12.2% and ≤16.4%, respectively. The assay was linear from 0.14 to 53.5 pM and 1.1 to 91.6 pM, with a lower limit of the measuring interval of 2.8 and 8.8 pM for RGFFYTPKTRREAE and GSLQKRGIVE, respectively. The concentration of des-31,32-proinsulin increased more during childhood than intact proinsulin. Participants with type 1 diabetes-associated autoantibodies had higher proinsulin-to-C-peptide ratios. Method comparison with 2 commercial immunoassays revealed variable cross-reactivity with insulin.
Conclusion:
The validated assay is robust and will be a useful tool for advancing studies of β-cell function. A detailed standard operating procedure, well-characterized calibration material, and monoclonal antibodies are available for adoption in other laboratories.
More Related Videos
06:09Quantification of Proteins Using Peptide Immunoaffinity Enrichment Coupled with Mass Spectrometry
Published on: July 31, 2011
08:07Identification of Inositol Phosphate or Phosphoinositide Interacting Proteins by Affinity Chromatography Coupled to Western Blot or Mass Spectrometry
Published on: July 26, 2019
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Production of Pharmaceuticals