Related Experiment Video
Updated: Aug 25, 2026

Surgical Treatment for Benign Prostatic Hyperplasia: Holmium Laser Enucleation of the Prostate (HoLEP).
Published on: March 6, 2018
Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser
Jessica Wu1, Amanda E Kahn2, Benjamin Dalkin2
1School of Medicine, University of Rochester, Rochester, New York, USA.
Introduction:
Operating room (OR) noise during holmium laser enucleation of the prostate (HoLEP), particularly from laser activation, may impair surgical workflow and staff performance.
Methods:
84 HoLEP procedures were recorded at three academic institutions between June 2024 and January 2026. Laser platforms included Lumenis Moses 2.0, Quanta Litho 150, Quanta Magneto, and Dornier Thulio. Sound measurements were obtained using a standardized smartphone-based system during three phases: laser-off (baseline), idle (not firing), and active lasing. Acoustic metrics included average sound level (LAeq), peak sound level (Lcpeak), and sound levels exceeded for 10%, 50%, and 90% of the recording time (L10, L50, and L90). Background noise correction was performed using baseline measurements to account for ambient noise and isolate laser-related sound. OR staff completed a validated survey assessing perceived noise burden (0-10 scale).
Results:
Background-corrected noise during active lasing differed significantly by platform. Litho demonstrated lower mean noise levels compared to Thulio for LAeq (60.6 vs. 68.2 dB, p = 0.023) and L10 (61.5 vs. 70.6 dB, p = 0.023), while Moses had higher LAeq compared to Litho (63.1 vs. 60.6 dB, p = 0.043). Overall, Litho had the lowest noise levels, whereas Thulio had the highest. Survey responses paralleled these findings. Litho was associated with lower perceived overall effort (2.1 ± 0.9), while Moses demonstrated higher overall effort (3.6 ± 1.2) for communication and concentration (Figure 2). Significant pairwise differences were observed between Moses and Litho (all p < 0.05). Trends were consistent across sites.
Conclusions:
Noise levels during HoLEP varied by laser technology and correlated with differences in perceived communication and cognitive effort. Litho demonstrated the lowest acoustic burden, while Moses and Thulio had higher noise exposure. These findings highlight the impact of laser selection on the OR environment and intraoperative team performance.