Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery
Kaushik Gokul1, Rohesh Fernando2, Bryan Marchant2
1Department of Internal Medicine, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
Background:
Leadless pacemaker use has grown with some now including temperature-based sensors rather than motion-based sensors. We present a case of unexpected rewarming-induced rapid pacing during cardiopulmonary bypass in a patient with the Aveir leadless pacemaker.
Case Summary:
We present a case of a patient with a leadless pacemaker undergoing cardiac surgery and suffering from rapid ventricular pacing when weaning from bypass treated by applying a magnet to stop rapid pacing with device interrogation conducted intraoperatively. The rate response feature was disabled with the patient weaned from bypass.
Discussion:
Traditional pacemakers use rate response features whereas the Aveir leadless pacemaker uses a temperature-based response feature. For this reason, the device is susceptible to inappropriate pacing when patients undergo rewarming during cardiac surgery resulting in possible hemodynamic compromise.
Take-Home Message:
Patients undergoing cardiac surgery with leadless pacemakers in place should have a careful preoperative evaluation with the rate response feature turned off before surgery if indicated.