Purpose:
Standardization and international guidelines for stent size selection are lacking. In this study, we introduce and validate an artificial intelligence (AI)- and image processing-supported, modular software algorithm trained on a multicentric vascular segmentation dataset that identifies stenoses, performs segmentations of stenotic vessel segments and suggests the optimal stent size for implantation.
Material And Methods:
This retrospective multicenter study included 149 patients who underwent stent implantation for symptomatic stenoses of the common and external iliac arteries between August 2017 and July 2024. Peri-interventional angiography datasets were evaluated by four board-certified interventional radiologists. For AI-training, all relevant stenoses were annotated and segmented to reflect intended stent sizing. The segmentation criteria were consensus-defined, and all readers completed a prior training session to ensure consistency. The modular algorithm comprises components for stenosis detection, segmentation and stent parameter prediction. Following pre-training on a publicly available coronary artery dataset, the model was fine-tuned on the study-specific iliac artery dataset using leave-one-out cross-validation.
Results:
OSCAR detected stenoses in 84.6% of cases. The model achieved a high recall (0.89 ± 0.21), meaning that most expert-annotated stenoses were correctly identified, while a moderate precision (0.65 ± 0.28) indicated some false-positive detections. Segmentation accuracy was good (DSC 0.77 ± 0.11). Stent diameter and length predictions demonstrated mean absolute percentage errors of 0.13 ± 0.18 and 0.33 ± 0.31, respectively, comparable to expert variability.
Conclusions: