Loss of Biventricular Capture After CRT-D Implantation: Device Malfunction or Something Else?
Paul Weber1, Dhruva Govil2, Dipak Shah1
1Department of Cardiology, Henry Ford Health Providence, Southfield, Michigan, USA.
Case Summary:
A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction.
Take-Home Message:
Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias