Related Experiment Video
Updated: Sep 5, 2026

Single-port Robotic Transanal Total Mesorectal Excision in a Porcine Model (Sus scrofa domesticus)
Published on: April 3, 2026
A low-cost indigenous self-retaining perineal retractor: Design and early clinical evaluation
Kanhaiya Hariyale1, Yogesh Tiwari2, Utkarsh Ghormare3
1Senior Resident, Department of Surgical Oncology, State Cancer Institute, NSCB Medical College, Jabalpur, India.
Abstract:
Commercial self-retaining perineal retractors provide excellent exposure during anorectal and vaginal surgery but remain prohibitively expensive for many resource-limited hospitals. We designed an indigenous self-retaining perineal retractor using readily available materials, incorporating a re-usable stainless-steel ring, modified barbless fishing hooks, disposable Foley catheter traction strings and a novel self-locking slot mechanism. The device was prospectively evaluated in three patients undergoing colorectal and gynecological procedures. It provided stable self-retaining circumferential exposure, facilitated precise dissection and suturing and was not associated with device-related complications. Costing approximately US$10, this reusable system offers an affordable, versatile alternative for perineal surgery in low-resource settings.