Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
Fluorinated radiopharmaceuticals sorption onto plasticized polyvinyl chloride automated injector tubings:
Lola Kazek1, Perrine Tylski2, Anthime Flaus3
1Hospices Civils de Lyon, Hôpital Louis Pradel, service pharmaceutique, Radiopharmacie, Bron, France.
Background:
The use of automated injectors for the administration of radiopharmaceuticals limits staff exposure and optimises the dose administered to patients, but many of these devices are only validated for a limited number of radiopharmaceuticals. The objectives of this study were to investigate the physicochemical phenomena occurring during the manipulation of fluorinated radiopharmaceuticals when in contact with injector tubings and to quantify and qualify the interactions between the radiopharmaceuticals and the tubing materials.
Results:
For six of the radiopharmaceuticals tested, the measured doses reached 97.64 ± 0.24% of the programmed dose. These results indicate an absence of sorption phenomena and validate their use on the injector. In contrast, the three most lipophilic molecules caused the injector to malfunction, resulting in significant residual activity in the tubing of up to 67%. These results were confirmed by activimeter measurements and positron emission tomography images.
Results:
from electronic microscopy results showed an alteration of the inner surface of the tubings that came into contact with lipophilic radiopharmaceuticals; however, this modification did not drastically alter the surface Zeta potential. Infrared spectroscopy confirmed the composition of the tubings as being plasticized polyvinyl chloride (PVCp), and plasticizer identification and quantification showed the presence of di-(2-ethylhexyl) adipate (reaching 37.5%) and di-(2-ethylhexyl) phthalate (more than 20%). A clear improvement was observed when co-extruded polyethylene / PVPp tubings were used.
Conclusions:
Unlike lipophilic radiopharmaceuticals, which exhibit sorption behaviour dependent on the presence of plasticizers, the tested hydrophilic radiopharmaceuticals demonstrated minimal interaction with PVCp tubing and can be reliably administered using the automated injector.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography
Drug Accumulation During Multiple Dosing: Repetitive IV Injections

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)