Comparative evaluation of inductively coupled plasma mass spectrometry and a phosphatase method for lithium
1Beijing Key Laboratory of Intelligent Drug Research and Development for Mental Disorders, National Clinical Research Center for Mental Disorders, National Center for Mental Disorders, Beijing Anding Hospital, Capital Medical University, Beijing, China.
Abstract:
The objective of this study was to compare lithium results obtained by inductively coupled plasma mass spectrometry (ICP-MS) and a phosphatase method in paired patient serum samples and to summarize the available analytical evidence from aqueous-standard verification and routine quality-control records. A total of 171 paired serum samples were included in all method-comparison analyses, comprising 141 non-hemolyzed/non-chylous samples, 17 hemolyzed samples, and 13 chylous samples. ICP-MS analytical response was assessed using traceable aqueous lithium standards, while imprecision was summarized descriptively from routine quality-control records. Matrix-matched trueness verification, spike-recovery, standard-addition, and other serum-specific validation procedures were not performed. Although the two methods showed a strong linear association (r = 0.989), Deming regression demonstrated a constant positive bias of the phosphatase method relative to ICP-MS. At the four predefined concentrations of 0.5, 1.0, 1.2, and 2.0 mmol/L, the predicted phosphatase-method results exceeded the upper bounds of the corresponding acceptable predicted-result intervals. Manufacturer-reported LOD and LOQ values were lower for ICP-MS, but study-specific verification was not performed. These findings identify clinically relevant between-method bias under the analytical conditions of this study but do not constitute full matrix-specific validation of ICP-MS for serum lithium measurement. Prospective matrix-matched validation is required before broader clinical implementation.
More Related Videos
09:22Absolute Quantitation of Inositol Pyrophosphates by Capillary Electrophoresis Electrospray Ionization Mass Spectrometry
Published on: August 13, 2021
07:38Functional Characterization of Na+/H+ Exchangers of Intracellular Compartments Using Proton-killing Selection to Express Them at the Plasma Membrane
Published on: March 30, 2015
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.