Impact of Total Protein and Triglycerides on the comparability of Sodium and Potassium Measurements by Direct and
Parul Chopra1, Ginni Bharti2, Ankita Gupta1
1Department of Laboratory Medicine, Delhi State Cancer Institute, New Delhi.
Background:
Electrolyte analysers use ion selective electrode (ISE) technologies, either 'direct'(dISE) or 'indirect'(iISE). Indirect methods are susceptible to errors in samples with high protein or lipid content. This study evaluated the individual and combined effects of elevated total proteins(TP) and triglycerides(TG) on sodium(Na+) and potassium(K+) measurements.
Methods:
Fifty serum samples were pooled to prepare a low-protein, low-lipid matrix. Three enrichment series were generated using graded additions of 20% human albumin, 20% Intralipid, or both to achieve varying TP and TG. All samples were analysed for Na+ and K+ using dISE and iISE. The difference between methods (iISE-dISE) was correlated with TP and TG using regression and Bland-Altman analyses. A multivariable model assessed predictors of electrolyte bias.
Results:
Albumin addition produced negative sodium bias, reaching upto 10% at TP>13g/dL, while potassium changes remained small. Triglycerides caused marked sodium underestimation at high levels (>2000 mg/dL) with minimal effect on potassium levels. When both interferents were elevated, sodium disagreement was high (bias -5.4mmol/L; limits of agreement -19.1 to +8.2), reflecting a strong cumulative effect. Multivariable regression showed TP as the strongest predictor of Na+ bias (R2 = 0.67), with minimal contribution by TG.
Conclusion:
This experimental study demonstrates that increasing concentrations of total protein and triglycerides, particularly when both are elevated produce significant negative biases in sodium measured by iISE, while potassium values remain comparable. These findings become clinically relevant in critical care/ cancer patients where they receive albumin and/or lipid infusions and dISE/iISE results are used interchangeably for patient monitoring.
More Related Videos
11:51Application of Electrophysiology Measurement to Study the Activity of Electro-Neutral Transporters
Published on: February 3, 2018
12:48Measuring Cation Transport by Na,K- and H,K-ATPase in Xenopus Oocytes by Atomic Absorption Spectrophotometry: An Alternative to Radioisotope Assays
Published on: February 19, 2013
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Potentiometry: Membrane Electrodes
Ionic Strength: Effects on Chemical Equilibria
In this solution, the primary cation—the calcium...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...