Sensor-Driven Pacemaker Tachycardia During Chest Physiotherapy: The Hidden Accelerometer Sensor Response in a
Chu-Pak Lau1, Kathy Lai-Fun Lee1, Hung-Fat Tse1,2,3
1Cardiology Division, Department of Medicine, School of Clinical Medicine, Queen Mary Hospital, the University of Hong Kong, Hong Kong, SAR, China.
Introduction:
Leadless pacemakers (LPM) provide ventricular pacing, atrio-ventricular (AV) synchrony and rate response.
Background:
In the VDD mode, the single-chamber LPM is capable of atrial sensing with additional accelerometer activity sensing rate response.
Methods And Results:
An 85-year-old man with tachy-bradycardia received a Micra AV2 (Medtronic Inc). During chest physiotherapy, pulse rate increased to over 100 bpm during chest percussion when performed on the left side. This was due to VDD mode switching to VVIR pacing when the accelerometer sensor rate exceeded the atrial rate. Inactivation of mode switching prevented rate acceleration.
Conclusion:
Chest physiotherapy caused unexpected rate increase in LPM even in VDD mode.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Dysrhythmias III: Characteristics of Dysrhythmias
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Holter Monitor: 24-Hour Monitoring