Jaka Strel1, Metka Križič1, Biljana Nikolova Stoilova1
1Arcus Medici Family Medicine Clinic, Žiri.
Objective:
The objective of this study was to clinically validate the MESI mTABLET BP device in accordance with the protocol outlined in ISO 81060-2:2018 + A1 + A2.
Methods:
A total of 127 adult participants were enrolled, of whom 36 were excluded. The final study sample consisted of 91 participants with upper arm circumferences ranging from 22 to 50 cm. Blood pressure measurements obtained with the MESI mTABLET BP were compared with reference auscultatory measurements performed simultaneously by two trained observers. A same-arm sequential method was used, in which each MESI mTABLET BP measurement was compared against the average of the preceding and subsequent reference measurements. The study was approved by the National Medical Ethics Committee of the Republic of Slovenia.
Results:
The mean blood pressure difference between MESI mTABLET BP and the reference auscultatory method was 1.6 ± 4.3 mmHg for SBP and 1.9 ± 5.1 mmHg for DBP; hence, criterion 1 from the standard was fulfilled (required |mean| ≤5.00, SD ≤8.00 mmHg). Criterion 2 was also met, as the SD of average paired determinations per participant was ±3.56 mmHg for SBP and ±4.66 mmHg for DBP (required ≤6.76 and ≤6.68 mmHg, respectively).
Conclusion:
The MESI mTABLET BP demonstrated clinical accuracy and reliability, and can hence be recommended for use in clinical and primary care settings in adult populations with upper arm circumferences between 22 and 50 cm, as validated in the present study.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure