Related Experiment Video
Updated: Sep 19, 2026

An Automated Culture System for Use in Preclinical Testing of Host-Directed Therapies for Tuberculosis
Published on: August 16, 2021
Optimization of culture read times using full microbiology laboratory automation
Janiece S Glover1, Bailey B Demarest1, Pam J Foster1
1Department of Pathology, Microbiology, and Immunology, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Abstract:
Clinical microbiology laboratories are increasingly adopting automation to meet testing demands, improve throughput, and accelerate diagnostic turnaround. This study evaluated optimal culture read times using Copan WaspLab microbiology laboratory automation (MLA) to determine if earlier read times affect accuracy and turnaround times as they pertain to microbial identification and antimicrobial susceptibility testing (AST). A total of 374 patient specimens were cultured simultaneously using a conventional incubation method and MLA, with MLA images captured from 10 to 72 h of incubation. Non-urine specimens were ready for workup in over 50% of cases at 16 h, 92% by 24 h, and 100% by 36 h, while 90% of urine cultures were ready for workup by 16 h. Using frozen isolates (127 gram-positive bacteria, 83 gram-negative bacteria, and 12 yeasts), we further evaluated the accuracy of MALDI-TOF identification and AST results for cultures incubated for 10 h and 16 h in MLA. Identification accuracy remained high across all groups, with gram-negative bacteria achieving 98.8% accuracy at 10 h and 97.6% at 16 h, gram-positive bacteria showing 97.6% accuracy at 10 h and 98.4% at 16 h, and yeasts demonstrating 91.6% accuracy at 10 h and 100% at 16 h of incubation, respectively. AST results using cultures at both time points demonstrated essential agreement above 90% for most drugs on Vitek 2 GN-807N, GP75, and ST02 AST cards. Minor and major errors were rare; however, elevated error rates were observed for specific drugs (e.g., cefazolin, piperacillin/tazobactam, nitrofurantoin, and linezolid), largely due to a limited number of susceptible isolates and breakpoint-related challenges. MLA enables earlier culture reads while maintaining ID and AST accuracy.
Importance:
Microbiology laboratory automation (MLA) is increasingly used to address rising test volumes and workforce limitations; however, optimal incubation durations for automated workflows are not well defined. In this study, we demonstrate that cultures from a range of specimen types can be reliably read as early as 16 h and finalized by 36 h without compromising organism identification or antimicrobial susceptibility testing accuracy compared to conventional incubation. These findings support the use of earlier culture read time points in MLA workflows, enabling faster turnaround of clinically actionable results. Implementation of optimized incubation times has the potential to improve laboratory efficiency and accelerate patient care decisions while maintaining diagnostic reliability.
More Related Videos
06:24Generic Protocol for Optimization of Heterologous Protein Production Using Automated Microbioreactor Technology
Published on: December 15, 2017
08:58Efficient Sampling of Genetically Encoded Biosensor Design Space Enabled with a Design of Experiments and Automation Workflow
Published on: October 17, 2025
Related Concept Videos
Automated Microbial Diagnostics
Key Techniques in Microbiology